4.6 Main board and Power Supply
Warning messages
ActionPossible causeDescriptionWarning code
Tap Continue on the Touch
screen.
If the system is in a cold
chamber, make sure that the
temperature is not below
+4°C.
Make sure that there is ade-
quate ventilation surround-
ing the system.
Make sure that the system
is not standing in direct sun
light.
System operation is not af-
fected by this warning, the
system can continue to be
used.
Instrument temperature is out
of range (4°C to 35°C)
System over-
temperature
011
Error messages
ActionPossible causeDescriptionError code
Contact a GE Service Engi-
neer.
EPROM error001
Contact a GE Service Engi-
neer.
The system cannot be used
in this condition.
Instrument temperature is out
of range (4°C to 35°C)
MPWAtemper-
ature
002
Contact a GE Service Engi-
neer.
With assistance from a GE
Service Engineer, check the
Power supply for voltages.
Voltage out of operating range.Power supply
monitor error
003
100 ÄKTA start Maintenance Manual 29060308 AD
4 Troubleshooting
4.6 Main board and Power Supply