3.6.2 Diagnostics
Parameter description
DescriptionParameter
Setting of the desired flow rate (ml/min).Flow rate
Displays the actual number of run hours for the Pump.Pump run
Check the Pump flow rate
Follow the instructions below to perform Pump diagnostics.
ActionStep
Immerse the buffer inlet tubing A in demineralized water.1
In the Pump screen, tap Diagnostics.2
Result: The Diagnostics screen opens.
Use the up/down arrows to set the desired Flow Rate value in the range 0.5
to 5 ml/min.
3
52 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.6 Pump
3.6.2 Diagnostics