3.8 Sample valve
Description
The Sample valve can be switched to allow the inlet of either buffer or sample.
DescriptionParameter
Indicates how many times the valve has switched.
When a valve has been replaced, the number of counts
has to be reset to 0.
Turn valve switches the valve between the Buffer and
Sample inlet ports. This option is used for troubleshoot-
ing the valve.
Valve switches
Instructions
Follow the instructions below to check if the Sample valve functions properly.
ActionStep
Immerse the buffer inlet tubing A and the sample inlet tubing in deminer-
alized water.
Place the outlet tubing from the Wash valve in the waste bottle.
Check that Buffer A is selected in the Buffer valve screen.
1
Start the Pump:2
In the Settings and service screen, access Pump:Diagnostics screen.
Enter the desired flow rate and then start the Pump by tapping Start
flow.
In the Settings and service screen, tap Next to access the 2
nd
screen.3
Result: The following screen opens.
ÄKTA start Maintenance Manual 29060308 AD 59