3.10 Outlet valve
Description
The Outlet valve can be switched to divert the flow, either to waste or to the Fraction
collector.
DescriptionParameter
Indicates how many times the valve has switched.
When a valve has been replaced, the number of counts
has to be reset to 0.
Turn valve switches the valve between the Waste and
Collection outlet ports. This option is used for trou-
bleshooting the valve.
Valve switches
Instructions
Follow the instructions below to check if the Outlet valve functions properly.
ActionStep
Immerse the buffer inlet tubing A in demineralized water.
Place the outlet tubing from the Outlet valve in the waste bottle.
Check that:
-
Buffer A is selected in the Buffer valve screen
-
Buffer is selected in the Sample valve screen
-
Column is selected in the Wash valve screen
1
Start the Pump:2
In the Settings and service screen, access Pump:Diagnostics screen.
Enter the desired flow rate and then start the Pump, tap Start flow.
ÄKTA start Maintenance Manual 29060308 AD 65
3 Maintenance operations
3.10 Outlet valve