5.9 Injection valve kit
Introduction
Use the Injection valve kit to service the valve when the external parts and liquid distri-
bution parts are damaged, worn or blocked with salt deposits.
Replace the complete valve (see Section 5.8 Injection valve, on page 136) when internal
valve parts are damaged, worn or blocked with salt deposits.
Required tools
DimensionTool
T10Torx driver
Instructions
Follow the instructions below to remove and replace the Injection valve kit.
ActionStep
Make sure that the valve is in position: Inject, and then disconnect the inlet
and outlet tubing from the ports.
1
ÄKTA start Maintenance Manual 29060308 AD 139
5 Removing and replacing modules
5.9 Injection valve kit