3.13.2 Diagnostics
Parameter description
DescriptionParameter
Checks that Frac30 rotates the Bowl assembly correctly
and shifts one tube at a time.
Feed tube test
Checks that Frac30 rotates the Bowl assembly correctly
and shifts from the current position to the home position
(tube number 1).
Home test
Instructions
Follow the instructions below to run a diagnostics test on Frac30.
ActionStep
In the Fraction collector screen, tap Diagnostics.1
Result: The Diagnostics screen opens.
Tap Feed tube test and verify that Frac30 rotates from one tube to the next.2
ÄKTA start Maintenance Manual 29060308 AD 79
3 Maintenance operations
3.13 Frac30
3.13.2 Diagnostics