3.5.3 Flow cell path length
Description
A Flow cell path length test is used to derive the actual path length of the UV flow cell.
The test should be performed when the UV Monitor or the UV flow cell has been replaced,
and when normalized UV absorption comparisons between different systems are
needed.
Parameter description
DescriptionParameter
Path length of the UV flow cell (mm).Cell
The absorbance of the liquid in the UV flow cell (mAU).Abs
UV LED drift compensation constant.C drft
Ambient temperature compensation constant.C amb
NOTICE
Do not change the C drft and C amb values. These parameters
must be set by a GE Service Engineer.
Prerequisites
Buffer A: immerse Buffer A inlet into demineralized water.
Buffer B: freshly prepared 1.0% acetone solution (vol/vol), which is expected to give
an absorbance value of 340 mAU.