3.6.3 Pump tubing log
Parameter description
DescriptionParameter
Indicates how many hours the pump tubing has been
used.
When the pump tubing is replaced, use the Reset but-
ton to reset the number of hours to 0.
Tubing run
Instructions
Follow the instructions below to check the Pump tubing log and to reset the number of
hours the pump tubing has been used.
ActionStep
Tap Pump tubing log.1
Result: The Pump tubing log screen opens.
In the Pump tubing log screen, tap Reset.2
Result: A confirmation screen opens.
54 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.6 Pump