4.5 Pressure sensor
Error messages
ActionPossible causeDescriptionError code
Check tubing and valves by
disconnecting one at a time
starting from the Fraction
collector going backwards
towards the Pump. When
pressure is released the
blockage has been found,
clean or replace tube to re-
solve the issue.
Clean the column/separa-
tion media with suitable solu-
tion (1 M NaOH) or replace
the column with a new one.
Check the specification and
pressure of the columns for
correct flow rate.
Blockage in the tubing,
valves or in the column.
Improperly selected Flow
rate.
Wrong pressure setting used
for chosen application.
Overpressure501
Perform Zero offset.
Contact a GE Service Engi-
neer.
The sensor is not connected or
not calibrated.
Pressure sen-
sor failure
502
ÄKTA start Maintenance Manual 29060308 AD 99
4 Troubleshooting
4.5 Pressure sensor