3.12 Pressure sensor
Parameter description
DescriptionParameter
Displays the current pressure in the flow path (MPa).P set
Instructions
Follow the instructions below to set the pressure to 0 when the Pressure sensor is ex-
posed to atmospheric pressure only.
NOTICE
Calibration of Pressure sensor must be performed by a GE Service
Engineer, and must be done when the Pressure sensor has been
replaced.
ActionStep
Disconnect the inlet tubing from the Pressure sensor to expose the sensor
to atmospheric pressure only.
Note:
Make sure that the Pump is OFF before disconnecting the tubing.
1