3.11.1 Temperature sensor calibration
Instructions
Follow the instructions below to calibrate the temperature sensor.
ActionStep
Place a precision thermometer in the flow cell path directly after the Con-
ductivity flow cell (i.e., immediately after the conductivity tubing which is
connected to Outlet valve), and then pump demineralized water through
the system with a flow rate of 0.5 ml/min.
1
In the Conductivity screen, tap Advanced calibration.2
Result: The Advanced calibration screen opens.
Note the temperature and enter it into the Set actual temp text box, and
then tap Calibrate to carry out the temperature calibration.
Note:
Make sure that the temperature of the Conductivity flow cell has stabilized
and measure the temperature of the calibration solution with a precision
thermometer.
3
ÄKTA start Maintenance Manual 29060308 AD 69
3 Maintenance operations
3.11 Conductivity Monitor
3.11.1 Temperature sensor calibration