3.14.2 Export system report to USB
Instructions
Follow the instructions below to export the system report to a USB memory stick. Use
the system report in further contacts with GE Service Engineers.
ActionStep
Plug the USB memory stick into the USB port located on ÄKTA start.1
In the System screen, tap Export System Report to USB.2
Result: The files INSTLOG.TXT and ERRORLOG.TXT are exported to the USB
memory stick.
Remove the USB memory stick from the USB port and connect it to a com-
puter.
3
Check the content of the system report files INSTLOG.TXT and ERRORLOG.TXT,
see section System report file parameters.
4
System report parameters
The content of the INSTLOG.TXT file contains the following parameters:
ÄKTA start serial no.
Firmware version
FPGA version
Pump run
Pump tube run
UV LED run time
Buffer valve count
Wash valve count
Outlet valve count
Sample valve count
The content of the ERRORLOG.TXT file contains Error codes, for example:
501: Over Pressure
301: Fraction collector failure
603: Illegal operation, restart instrument
88 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.14 System
3.14.2 Export system report to USB