ActionStep
Once the firmware is updated, perform the following calibration sequence:7
Display:Touch screen calibration
Pressure sensor:Zero offset
Pump : Calibration:Flow rate
UV:UV LED calibration
Conductivity:Advanced calibration (temperature sensor calibration)
Conductivity:Calibration (flow cell calibration)
ÄKTA start Maintenance Manual 29060308 AD 87
3 Maintenance operations
3.14 System
3.14.1 Firmware update