ActionStep
Tap Microprocessor firmware update.3
Result: The following screen opens.
Tap Yes and then wait for approximately 3 minutes for the update to be
done.
Note:
When the firmware update is completed, the instrument automatically restarts
and displays the version of the firmware.
From the product support page (www.gelifesciences.com/AKTAstart),
download AKTASTRT.dat for FPGA update.
4
Tap FPGA update.5
Result: The following screen opens.
Tap Yes. The updating may take up to 5 minutes.6
Result: The following screen opens.
86 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.14 System
3.14.1 Firmware update