3.7 Buffer valve
Description
The Buffer valve can be switched to allow the inlet of either buffer A or buffer B, or a
mixture of A and B (gradient).
DescriptionParameter
Indicates how many times the valve has switched be-
tween buffer A or buffer B.
When a valve has been replaced, the number of counts
has to be reset to 0.
Turn valve switches the valve between the A and B
inlet ports. This option is used for troubleshooting the
valve.
Valve switches
Instructions
Follow the instructions below to check if the Buffer valve functions properly.
ActionStep
Immerse the buffer inlet tubing A and B in demineralized water.
Place the outlet tubing from the Wash valve in the waste bottle.
1
Start the Pump:2
In the Settings and service screen, access Pump:Diagnostics screen.
Enter the desired flow rate and then start the Pump by tapping Start
flow, and then tap Back:Back to return to the Settings and service
screen.
56 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.7 Buffer valve