ActionStep
In the Settings and service screen, tap Next to access the 2
nd
screen.3
Result: The following screen opens.
In the Settings and service screen, tap Buffer valve.
Result: The Buffer valve screen opens.
Check that the valve position Buffer A is selected.
Tap Turn valve to check that the Buffer valve switches between Buffer
A and Buffer B on the Display. There is also a click sound when the valve
switches.
Check that the flow is delivered from the Buffer A inlet or the Buffer B
inlet, according to the selected valve position.
Note:
The Buffer valve is by default in the Buffer A position (the liquid is delivered
from the Buffer A inlet).
4
Visually inspect the flow at the outlet.5
If the water is not flowing properly, check the following and fix accordingly:
Is there a blockage in the tubing?
Are the connectors tightly connected?
Stop the Pump:6
In the Settings and service screen, access Pump:Diagnostics screen.
Stop the Pump by tapping Stop flow.
ÄKTA start Maintenance Manual 29060308 AD 57
3 Maintenance operations
3.7 Buffer valve