Troubleshooting Flow chart 3
ERROR 103
UV Monitor Failure
1. Switch off the instrument.
2. Disassemble the UV Monitor.
3. Check that the UV Monitor RJ45 cable is secure;
replug properly.
4. Switch on the instrument.
Error message
still displayed?
Proceed with your work.
Contact GE Service Engineer.
Yes
No
104 ÄKTA start Maintenance Manual 29060308 AD