Troubleshooting Flow chart 6
ERROR 201
Conductivity Monitor Failure
1. Switch off the instrument.
2. Disassemble the Conductivity Monitor.
3. Secure the Cable properly and reassemble
the Conductivity Monitor.
Switch on the instrument.
1. Replace the Conductivity Monitor.
2. Enter the new Cell Constant value.
3. Perform Conductivity calibration.
Error message still
displayed?
Proceed with your work.
Has calibration
passed?
Contact GE Service Engineer.
No
Yes
Yes
No
ÄKTA start Maintenance Manual 29060308 AD 107
4 Troubleshooting
4.8 Troubleshooting flow charts