Instructions
Follow the instructions below to remove and replace the valve.
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
ActionStep
Switch off the instrument by pressing the Power Switch to the O position.1
Remove all tubing from the ports.2
Loosen the retaining screw from the valve fitting using a T10 torx driver,
supplied with the equipment at delivery.
Note:
The screw can be located on different places on the top circumference of the
holder.
3
ÄKTA start Maintenance Manual 29060308 AD 111
5 Removing and replacing modules
5.1 3-Port valves