ActionStep
Remove the Pump.6
Note:
After removing the Pump, make sure that the pump cable is placed clear of
the fan, on the left side of the instrument. This is to ensure that the cable does
not get stuck in the ventilation fan.
Replace with a new Pump, install the new Pump in reverse order.7
Place the Pump tubing between the rollers and the track of the new Pump.8
Switch on the instrument by pressing the Power Switch to the l position.9
In the Pump screen, tap Diagnostics and then reset the number of hours
of Pump run to 0.
10
128 ÄKTA start Maintenance Manual 29060308 AD
5 Removing and replacing modules
5.5 Pump