ActionStep
Loosen the screws from the Conductivity Monitor using a T20 screwdriver,
supplied with the equipment at delivery.
3
Slowly remove the Conductivity Monitor until a cable is visible. The cable is
assembled with two connectors interconnected.
Note:
If required, twist the connector to bring the connector out of the instrument
hole, before disassembly the Conductivity Monitor.
4
ÄKTA start Maintenance Manual 29060308 AD 133
5 Removing and replacing modules
5.7 Conductivity Monitor