ActionStep
Tap Yes to automatically adjust the Light strength value to get a minimum
Signal response of 2700 mV.
5
Tap Save.6
Result: The calibrated Light strength value is saved and stored in the perma-
nent memory on main board.
40 ÄKTA start Maintenance Manual 29060308 AD
3 Maintenance operations
3.5 UV Monitor
3.5.1 UV LED calibration