ConstraintsConnectorExternal module
Fraction collector F9-R, 2nd requires Fraction collector
F9-R or Fraction collector F9-C.
F-typeFraction collector F9-R, 2nd
Cannot be used at the same time as Mixer valve V9-M
or V9H-M.
D-typeSample pump S9 and S9H
To optimize signal quality, the total cable length connecting all external modules to the ÄKTA
pure instrument should not exceed 10 m.
Note:
ÄKTA pure connector plate
The connector plate with the UniNet-9 connectors is located on the back of the ÄKTA pure instrument.
9
FunctionConnector nameConnector
Connector point for service.Test9
NOTICE
Do not connect any module to the connector Test on the ÄKTA pure instrument.
Plug all unused UniNet-9 ports (ports 1 to 8) on the ÄKTA pure instrument with jumpers.
Note:
The connector Test should be protected by a plastic lid. Do not plug the connector with a
jumper.
Note:
ÄKTA pure User Manual 29119969 AB 123
3 ÄKTA pure external modules
3.6 Connection of external modules