9.8.2 System settings - Conductivity
The table below describes the conductivity related system settings available for ÄKTA pure.
DescriptionInstruction name
Alarm conductivity enables or disables the conductivity alarm.
When enabled, it sets the alarm limits for the conductivity signal.
When the conductivity falls outside the set limits, an alarm will be
triggered and the method will be paused.
Alarm conductivity
Relative scale cond facilitates monitoring of a gradient, for which
the user sets the conductivity values for 0% and 100%. The Rela-
tive scale cond can be set in ascending manner (0% for low and
100% for high conductivity) or in descending manner (0% for high
and 100% for low conductivity).
Note:
The Relative scale cond in descending manner is especially useful
for conductivity visualization in RPC and HIC, where the conductiv-
ity curve is reversed compared to the concentration curve (i.e., high
conductivity at 0% B and low conductivity at 100% B).
Relative scale cond
Cond temp compensation is used to adjust the conductivity values
to a reference temperature in order to compare conductivity val-
ues between runs that have been performed at different temper-
atures.
Cond temp condensation
Setting the compensation factor to 0% turns this function off.
ÄKTA pure User Manual 29119969 AB 451
9 Reference information
9.8 System settings
9.8.2 System settings - Conductivity