6.3 Air sensor B and Inlet valve B tests
Air sensor B and Inlet valve B tests
The following Air sensor B and Inlet valve B tests are available, for ÄKTA pure 25 and ÄKTA pure 150,
respectively:
Air sensor B and inlet valve V9-IB test
Air sensor B and inlet valve V9H-IB test
Method description
The Air sensor B and inlet valve B test switches the inlet valve B, labeled V9-IB or V9H-IB, to an empty
position and checks if the air sensor detects air.
The method run takes approximately 1 minute.
Required configuration
A correctly installed Inlet valve B is required to run the test.
Required material
The following materials are required:
Syringe, 25-30 ml
Distilled water
Prepare the test
Follow the instructions below to prepare the system before method start.
ActionStep
Direct the outlet tubing marked W to a waste container.1
Place inlet tubing B1 into distilled water.2
Prime the B1 inlet and purge the pump. See Section 5.4.1 System pumps, on page 162.3
Disconnect the inlet tubing connected to inlet valve position B3. During the test method
air is introduced into the inlet valve through that inlet port to test the function of the air
sensor.
4
ÄKTA pure User Manual 29119969 AB 231
6 Performance tests
6.3 Air sensor B and Inlet valve B tests