7.8.2 Replace internal tubing in Fraction collector F9-C
Maintenance interval
Replace tubing and connectors when required, for example when the tubing has clogged or has been
bent so that the flow is stopped.
Required material
The following material are required:
Tubing and connectors
Tubing cutter
Fingertight wrench
Instructions
Follow the instructions below to replace tubing and connectors. See Section 3.2.2 Fraction collector
F9-C illustrations, on page 93 for the locations of the components of the fraction collector.
ActionStep
Switch off the power to the fraction collector by switching off power to the ÄKTA pure
instrument.
1
Remove the connector for the internal tubing that is attached to the inlet port:2
Unscrew the connector from the inlet port.
Remove the connector from the tubing.
ÄKTA pure User Manual 29119969 AB 319
7 Maintenance
7.8 Replacement procedures
7.8.2 Replace internal tubing in Fraction collector F9-C