ActionStep
Remove the tubing connector from the dispenser head:3
Open the fraction collector and gently move the fractionation arm forward.
Unscrew the connector from the dispenser head and remove the tubing.
Gently move the fractionation arm all the way back.
Cut the new tubing to the same length as the old tubing. Use a tubing cutter to get a
straight angle cut.
4
CAUTION
Cut injuries. The tubing cutter is very sharp and must be handled with
care to avoid injuries.
Note:
Use tubing with the original inner diameter and length to ensure that the correct delay
volume is maintained.
320 ÄKTA pure User Manual 29119969 AB