7.3.1 Change pump rinsing solution
Maintenance interval
Replace the pump rinsing solution in the system pumps and the sample pump (optional module) every
week to prevent bacterial growth.
Required material
The following material are required:
20% ethanol
Syringe, 25-30 ml
Illustration of the system pump piston rinsing
system
The illustrations below show the parts and tubing of the system pump piston rinsing system
1
2
3
4
5
DescriptionPart
Rinsing system tube holder, top1
Rinsing system tube2
Rinsing system tube holder, bottom3
Outlet tubing4
Inlet tubing5
262 ÄKTA pure User Manual 29119969 AB
7 Maintenance
7.3 Weekly maintenance
7.3.1 Change pump rinsing solution