4.2.2 Configuration of Mixer valves
System properties
Follow the instruction below to update the system properties.
ActionStep
Open the system properties Edit dialog.1
Select Valves and pumps from the Component types list.2
Select Mixer valve (V9-M) or Mixer valve (V9H-M) in the Component selection list.3
System settings
The flow rate for Mixer by-pass wash can be set.
DescriptionInstruction name
Sets the flow rate used during Mixer by-pass wash and defines
wash volumes for mixer by-pass wash options.
Note:
The flow rate should not exceed 10 ml/min if narrow inlet tubing
(i.d. 0.75 mm) is used.
Mixer by-pass wash settings
ÄKTA pure User Manual 29119969 AB 133
4 System configuration
4.2 Configure modules
4.2.2 Configuration of Mixer valves

Table of Contents