EasyManua.ls Logo

GE AKTApure - Mixer

GE AKTApure
522 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.4.2 Mixer
Function of the Mixer
Mixer M9 is located after System pump A and System pump B, and before the Injection valve. The
Mixer is a dynamic mixer for high-performance gradients. It is used to make sure that the buffers from
the System pumps are mixed to give a homogenous buffer composition.
The Mixer has a built-in filter that prevents particles from reaching the column.
Replace the sealing ring of the mixer with a highly resistant O-ring if the system is going to be
exposed to organic solvents or high concentrations of organic acids for longer periods of time.
Refer to Section 7.8.4 Replace the O-ring inside the Mixer, on page 323 for how to change the
ring.
Note:
Location and illustration
The illustration below shows the location, together with a detailed view of Mixer M9.
1
2
3
DescriptionPart
Outlet1
Mixer chamber (0.6, 1.4, 5 or 15 ml)2
CAUTION
Risk of explosion. Do not use Mixer chamber 15 ml with
an ÄKTA pure 25 configuration. The maximum pressure
for Mixer chamber 15 ml is 5 MPa.
Inlet3
38 ÄKTA pure User Manual 29119969 AB
2 The ÄKTA pure instrument
2.4 Instrument modules
2.4.2 Mixer

Table of Contents

Related product manuals