9.3 Tubing and connectors
Tubing types
The table below shows the tubing types used in ÄKTA pure.
Volume/cmScope of useColorDescription
0.5 μlHigh pressure tubing
Reference capillary 1
Tubing Kit 0.25
BluePEEK, o.d. 1/16", i.d. 0.25
mm
2.0 μlHigh pressure tubing
Tubing kit 0.5 (stan-
dard)
OrangePEEK, o.d. 1/16", i.d. 0.50
mm
4.4 μlHigh pressure tubing
Tubing kit 0.75
GreenPEEK, o.d. 1/16", i.d. 0.75
mm
7.8 μlHigh pressure tubing
Tubing kit 1.0
BeigePEEK, o.d. 1/16", i.d. 1.0 mm
20.0 μlInlet tubingTransparentFEP, o.d. 1/8", i.d. 1.6 mm
66.0 μlInlet tubing for high
flow rate and high vis-
cosity
TransparentFEP, o.d. 3/16", i.d. 2.9 mm
4.4 μlNarrow inlet tubing
(optional)
TransparentETFE, o.d. 1/16", i.d. 0.75
mm
7.8 μlOutlet and waste tub-
ing
TransparentETFE, o.d. 1/16", i.d. 1.0 mm
0.3 mlWaste tubing from
Buffer tray
TransparentSilicone, o.d. 12 mm, i.d. 8
mm
Pump rinse solution
tubing
Note:
Many different sizes/types of tubing can be connected to a chromatography system. Tubing
with a smaller inner diameter (i.d.) holds less delay volume and will therefore generate less
dilution of the protein peak. Narrow tubing, however, increases the system pressure, espe-
cially when running at high flow rates. The tubing used should match the application needs.
See Section 9.4 Recommended tubing kits for prepacked columns, on page 428 for more
information.
When using the high pressure tubing kit with i.d. 1.0 mm to allow high flow rates in combi-
nation with high viscosities in the pumps, inlet tubing with a larger i.d. than standard might
be needed to avoid outgassing.
420 ÄKTA pure User Manual 29119969 AB