8.7 Troubleshooting: Other components
In this section
General Hardware: All modules
Mixer
Superloop
Cabinet
Power and ICU
External air sensors
I/O-box E9
Instrument control panel
General hardware: All modules
Possible cause and actionProblem
The cable between a module and the ICU is not connected
Remove the module and make sure that the cable is connected.
Two similar modules have been added to the instrument, for
example two Inlet valve A
Remove one of the modules with the same Node ID.
The Node ID for one or more of the modules is incorrect, for
example an Inlet valve A2 has the same Node ID as Inlet valve
A. The instrument then considers them to have the same
identity.
Remove Inlet valve A2 and change the Node ID according to
Section 9.16 Node IDs, on page 503.
Modules cannot be found by
the instrument
The Node ID for one or more of the modules is incorrect
Check Node ID and change the Node ID according to Section 9.16
Node IDs, on page 503.
An unknown instrument
module is connected to the
system
394 ÄKTA pure User Manual 29119969 AB
8 Troubleshooting
8.7 Troubleshooting: Other components