8.4 Troubleshooting: Valves
General
The following table lists the general problems that may occur for the different valves.
Possible cause and actionProblem
The cable between the valve and the ICU is not connected
Remove the valve and make sure that the cable is connected.
See Hardware installation of a module, on page 77.
Wrong Node ID
Check the valve Node ID. If necessary, change the Node ID. See
Node ID, on page 77.
The valve is not found by
the instrument
The air intake on the rear or on the left side of the instrument
is covered
Make sure that none of the air intakes on the instrument are
covered.
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal valve tempera-
ture is too high
Hardware error
Generate a System error report and contact Service.
The valve is not switching
or is switching to wrong po-
sition
Hardware error
Generate a System error report and contact Service.
External leakage
Broken valve
Replace the valve. See Section 2.5 Installation of internal modules,
on page 77.
Internal leakage
ÄKTA pure User Manual 29119969 AB 375
8 Troubleshooting
8.4 Troubleshooting: Valves