EasyManua.ls Logo

GE AKTApure - Page 263

GE AKTApure
522 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prime the system pump piston rinsing system
Follow the instructions below to fill the pump piston rinsing system with rinsing solution. See the tubing
configuration of the rinsing system in the illustration above.
ActionStep
Remove the rinsing system tube from the holder.1
Fill the rinsing system tube with 50 ml of 20% ethanol.2
Place the rinsing solution tube back in the holder.3
Insert the inlet tubing to the System pump piston rinsing system into the fluid in the
rinsing solution tube.
Note:
Make sure that the inlet tubing reaches close to the bottom of the rinsing solution tube.
4
Connect a 25 to 30 ml syringe to the outlet tubing of the System pump piston rinsing
system. Draw liquid slowly into the syringe.
5
Disconnect the syringe and discard its contents.6
Insert the outlet tubing into the fluid in the rinsing solution tube.7
Fill the rinsing solution tube so that the tube contains 50 ml of 20% ethanol.8
ÄKTA pure User Manual 29119969 AB 263
7 Maintenance
7.3 Weekly maintenance
7.3.1 Change pump rinsing solution

Table of Contents