ConstraintsRecommended locationModule
pH valve shall always be connected after
the column due to pressure constraints
and fractionation control. The delay vol-
ume calculations will be effected if the
valve is located elsewhere. Normally it is
situated between the monitor and the
Outlet valve.
2pH valve V9-pH or V9H-pH
Versatile valve V9-V or V9H-V has no
constraints.
Versatile valve 2 requires Versatile
valve V9-V.
Versatile valve 3 requires Versatile
valve 2.
Versatile valve 4 requires Versatile
valve 3.
NoneVersatile valve V9-V or
V9H-V
Versatile valve 2
Versatile valve 3
Versatile valve 4
UV monitor U9-M cannot be used at the
same time as UV monitor U9-L.
Note:
UV monitor U9-L, 2nd can be used at the
same time as UV monitor U9-M.
10UV monitor U9-M
UV monitor U9-L:11UV monitor U9-L
Cannot be used at the same time as
UV monitor U9-M.
Requires the Multi-module panel.
Note:
UV monitor U9-L, 2nd can be used at the
same time as UV monitor U9-L.
UV monitor U9-L, 2nd requires UV monitor
U9-L or UV monitor U9-M.
NoneUV monitor U9-L, 2nd
ÄKTA pure User Manual 29119969 AB 127
4 System configuration
4.1 Configuration overview