ActionStep
Select a system.
Select Control mode.
Click OK.
Result: The selected instrument can now be controlled by the software.
2
Select Tubing and Delay Volumes and select Tubing: Injection valve to column.
Select the I. D. from the drop-down list.
Type in the length of the tube in the Length field and click OK.
3
Perform step 3 for the instruction Tubing: Sample pump to injection valve.4
The built-in pressure sensors for Column valves V9-C and V9H-C have to be re-calibrated after
installation. See Calibrate the monitors, on page 305.
Note:
A Pre-Column pressure alarm shall always be set to protect the column, see Section 5.6 Pressure
alarms, on page 177. Column valve V9-Cs and V9H-Cs does not contain pressure monitors.
See Section 5.6 Pressure alarms, on page177 for how to protect columns when not using Column
valves V9-C and V9H-C.