ActionStep
In the Maintenance Manager dialog box, click New System Notification.1
Result: The NewNotification field appears in the Maintenance Manager dialog box.
In the NewNotification field:2
Enter a name for the new notification.
Select a time interval after which the new notification will be issued.
If desired, write a message that will be shown for the maintenance notification.
Click Apply to save the changes and apply the notification settings.3
Delete a user defined system notification
To delete a user defined system notification, select the notification in the Maintenance Manager dialog
box and press the Delete button.
Module notifications are predefined and cannot be deleted. If desired, they can be disabled.
Note:
Handle a maintenance notification
Each maintenance notification has a time interval after which the notification is issued. When this time
interval has been reached, a Maintenance Notification dialog box appears.
The possibility to add maintenance notifications for modules is only available for some systems,
but all systems can add maintenance notifications for the complete system.
Note:
Follow the instruction to handle the notification.
256 ÄKTA pure User Manual 29119969 AB