ActionStep
Fill a syringe with distilled water, and connect the syringe to the Luer connector.3
Squirt distilled water through the Conductivity flow cell in small amounts. Disconnect the
syringe.
4
Fill a syringe with 1 M NaOH, and connect the syringe to the Luer connector.5
Squirt 1 M NaOH through the Conductivity flow cell about five times.6
Leave the liquid in the flow cell for at least 15 minutes.7
Fill a syringe with distilled water. Connect the syringe to the Luer connector.8
Inject the distilled water into the Conductivity flow cell to rinse the flow cell. Disconnect
the syringe.
9
Disconnect the Luer connector from the top of the Conductivity flow cell, and reconnect
the fingertight connector with tubing.
10
ÄKTA pure User Manual 29119969 AB 301
7 Maintenance
7.6 Maintenance when required
7.6.10 Clean the Conductivity flow cell