Does the positioning of the columns correspond to the selections made in the method? For more
details, see Section 5.5 Connect a column, on page 173.
Purification checks
Have all columns been cleaned and prepared according to the column recommendations?
Have the samples been adjusted to binding buffer conditions?
Have the samples been clarified by centrifugation and/or filtration prior to sample loading?
Are the correct buffers used for the chosen columns and proteins?
Check buffers for precipitations. Adjust to room temperature.
Are the chosen columns suitable for the chosen target proteins?
Do the buffers have correct pH? The pH of some buffers changes with the temperature.
Are the UV-wavelengths used by the method appropriate with respect to used buffers and proteins?
For more details see the method handbooks available from GE.
ÄKTA pure User Manual 29119969 AB 359
8 Troubleshooting
8.2 Troubleshooting: General Checklist