Possible cause and actionProblem
The air intake on the rear or on the right side of the instrument
is covered
Make sure that none of the air intakes on the instrument are
covered.
Hot surroundings
Decrease the room temperature. Maximum operating tempera-
ture is 35°C.
Hardware error
Switch off the instrument and wait until the temperature has
decreased. Restart the instrument. If this error is recurrent, gen-
erate a System error report and contact Service.
The internal temperature of
the UV monitor is too high
Wrong UV flow cell for current buffer
Change to a shorter UV flow cell or change buffer.
The UV flow cell is not correctly installed
Check that the UV flow cell is fitted correctly, see Section 7.8.6
Replace the UV monitor U9-L flow cell, on page 328.
Dirt in the UV flow cell
Clean the UV flow cell, see Section 7.5.1 Clean the UV flow cell, on
page 273.
Broken UV flow cell
Replace the cell, see Section 7.8.6 Replace the UV monitor U9-L
flow cell, on page 328.
Maximum absorbance that
can be measured by the de-
tector is reached
Check that the UV flow cell is fitted correctly, see Section 7.8.6
Replace the UV monitor U9-L flow cell, on page 328.
The UV cell is not correctly
installed
ÄKTA pure User Manual 29119969 AB 365
8 Troubleshooting
8.3 Troubleshooting: Monitors