Possible cause and actionProblem
Air in the flow path
If this error is recurrent, generate a System error report and
contact Service.
Check the flow path for air. Fill system and purge pumps accord-
ing to Section 5.4.1 System pumps, on page 162.
Unclean Drop sync sensor diode windows
Clean the diode windows of the Drop sync sensor. See Sec-
tion 7.6.4 Clean Fraction collector F9-C, on page 288. If this error
is recurrent, generate a System error report and contact Service.
Too high flow rate
Decrease the flow rate.
The Dispenser head failed
to detect a drop
Air in the flow path
Check the flow path for air. Fill system and purge pumps accord-
ing to Section 5.4.1 System pumps, on page 162.
If this error is recurrent, generate a System error report and
contact Service.
Unclean diode windows in the Drop sync sensor
Clean the diode windows of the Drop sync sensor. See Sec-
tion 7.6.4 Clean Fraction collector F9-C, on page 288. If this error
is recurrent, generate a System error report and contact Service.
The Dispenser head failed
to detect the flow properly
and has switched to tube
change with reduced accu-
mulator functionality
Too high flow rate
Decrease the flow rate.
Air in the flow path
Check the flow path for air. Fill system and purge pumps accord-
ing to Section 5.4.1 System pumps, on page 162.
If this error is recurrent, generate a System error report and
contact Service.
Unclean diode windows in the Drop sync sensor
Clean the diode windows of the Drop sync sensor. See Sec-
tion 7.6.4 Clean Fraction collector F9-C, on page 288. If this error
is recurrent, generate a System error report and contact Service.
The Dispenser head failed
to detect a drop and has
switched to tube change
without Drop sync
Wrong tube dimensions are used
Check that the used tubes have the right dimensions. See Fraction
collector tubes and bottles, on page 100 for information about
tubes and Cassettes.
QuickRelease function is worn out
Order a new Cassette. See Chapter 10 for ordering information.
Tubes do not fit in the Cas-
sette
ÄKTA pure User Manual 29119969 AB 379
8 Troubleshooting
8.5 Troubleshooting: Fraction collector