Volume (μl)Length
(mm)
TubingDescriptionLabel
ÄKTA pure 150ÄKTA pure 25ÄKTA pure 150ÄKTA pure 25
310180400PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
Outlet valve to Frac-
tion collector F9-R
Frac
16070350PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
Outlet valve to Frac-
tion collector F9-C
Tubing of different inner diameter can be used for tubing Frac and tubing 5 to 9. An inner di-
ameter (i.d.) of 0.5 mm is standard, but tubing kits of 0.25 mm i.d., 0.75 mm i.d. and 1.0 mm
i.d. are also available. Tubing 3S is not included in the kits, but should be changed to 1.0 mm
i.d. if the Tubing kit 1.0 is used.
Note:
Optional high pressure tubing
Volume (μl)Length
(mm)
TubingDescriptionLabel
ÄKTA pure 150ÄKTA pure 25ÄKTA pure 150ÄKTA pure 25
13070160PEEK, o.d.
1/16", i.d. 1.0
mm
PEEK, o.d.
1/16", i.d. 0.75
mm
Pressure monitor to
Mixer valve
3-1
260150330PEEK, o.d.
1/16", i.d. 1.0
mm
PEEK, o.d.
1/16", i.d. 0.75
mm
Mixer valve to Mixer3-2
200120260PEEK, o.d.
1/16", i.d. 1.0
mm
PEEK, o.d.
1/16", i.d. 0.75
mm
Mixer valve to Injec-
tion valve
3-3
8040180PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
Conductivity monitor
to pH valve
8pH
7030160PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
pH valve to Outlet
valve
9pH
7030160PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
Injection valve port
LoopF to Loop valve
port F
L1
7030160PEEK, o.d.
1/16", i.d. 0.75
mm
PEEK, o.d.
1/16", i.d. 0.50
mm
Injection valve port
LoopE to Loop valve
port E
L2
ÄKTA pure User Manual 29119969 AB 425
9 Reference information
9.3 Tubing and connectors