DescriptionInstruction name
Alarm pre column pressure sets the alarm limits for the pre col-
umn pressure. When enabled and the pre column pressure falls
outside the set pressure limits, an alarm is issued and the method
will be paused. Default values for the alarm limits are set by the
values in the column list when a column is selected in the method
and Alarm pre column pressure is set as a variable. Low alarm
is only triggered if the pressure first exceeds the Low alarm limit
for ten seconds continuously then falls below the Low alarm limit.
Note:
Setting the Low alarm to 0 deactivates the low pressure alarm.
Alarm pre column pressure
454 ÄKTA pure User Manual 29119969 AB
9 Reference information
9.8 System settings
9.8.4 System settings - Pressure alarms