DescriptionInstruction name
Applicable if no fraction collector is used. Peak frac in outlet valve
enables collection of only those peaks that fulfill the conditions
set in Peak fractionation parameters. When the set fraction
size/outlet has been collected, the outlet valve turns to the next
position. A fraction mark is set in the chromatogram for each new
outlet position. If Outlet valve V9-Os or V9H-Os (1-outlet) is used,
only one peak can be collected via Outlet 1 position.
Peak frac in outlet valve
Applicable if no fraction collector is used. Stop peak frac in outlet
valve ends the peak fractionation in outlet valve after the set delay
volume (specified in System Settings:Tubing and Delay volumes)
has been collected. The outlet valve is then turned to position
Waste.
Stop peak frac in outlet valve
Peak fractionation parameters sets the detection parameters
for peak collection, i.e. it determines when a peak starts and ends.
This information is used by the instructions Peak fractionation,
Peak fractionation frac 2 and Peak frac in outlet valve in order
to start/end the peak collection.
Peak fractionation parameters
ÄKTA pure User Manual 29119969 AB 477
9 Reference information
9.9 Manual instructions
9.9.4 Manual instructions - Fraction collection