3 Checkout Procedure
M1046983 04/08 3-3
3.2.2 Machine Check The machine check runs automatically and beeps to indicate when it is
finished or if interaction is required.
The Machine Check does a:
• Machine Check - System check (Ventilator Circuit),
• Machine Check- Circuit check (Bag Circuit),
• and a Machine Check - Circuit O2 cell check (if circuit O
2
cell is present).
When one of the checks is completed, the system will transition to the next
check.
1. Turn the System switch to On.
2. Select Machine Check and follow the instructions.
3. If a check fails, follow the instructions to perform a recheck or accept the
results.
3.2.3 Machine Check -
System (Ventilator
Circuit Testing)
The Machine Check-System checks the Bag/Vent switch, proper gas supply
pressures, ventilator operation and leak, battery and electrical power, circuit
compliance, flow control operation, and vaporizer operation. This is a two-
step check.
Note The Machine Check-System can be run with a test cassette to invoke
extended diagnostics (Agent Delivery test will take several minutes longer).
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Select Start. The display shows the checks being run.
• The system beeps when this portion of the check is done.
• The results are shown on the display.
5. Make sure the bellows is fully collapsed.
6. Occlude the patient Y.
7. Select Continue. The display shows the checks being run.
8. When the check passes, the system will transition to the next step.
3.2.4 Machine Check -
Circuit (Bag Circuit
Testing)
The Machine Check-Circuit checks the Bag/Vent switch, proper gas supply
pressures, airway pressure measurement transducer, APL valve, and manual
circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.
4. (ACGO option only.) Set the ACGO switch to Circle.
5. Select Start. The display shows the checks being run.
• The system beeps when the check is done.
• The results are shown on the display.
6. When the check passes, the system will transition to the next step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters