7 Troubleshooting
M1046983 04/08 7-29
7.7 System Troubleshooting Flowcharts
Flowchart
6
AB.91.013
Start
Turn on System
Review Active Alarms, Error
Log and Alarm Log.
Proceed to "Display
Troubleshooting”
Yes
No
Does the System Application
Screen Appear?
Are there Alarms
relating to the
event?
Troubleshoot
Error/Alarm Logs
and Users Reference
Manual URM (Part 2)
Yes
Yes
Diagnosis
Achieved?
No
Proceed to "Breathing System
Leak Troubleshooting”
Proceed to "Inaccurate Volume
Ventilation Troubleshooting”
Proceed to "No Ventilation
Troubleshooting”
Yes
Yes
Yes
Yes
YesYes
No
No
No
No
Proceed to "High Intrinsic
PEEP Troubleshooting”
Yes
Yes
No
No
Alarms
Alarms of Type:
Breathing system loose
Circuit Leak
System Leak
Alarms of Type:
Calibrate flow sensors
Calibrate, Dry, or replace -
Flow Sensors
Check flow sensors
Replace insp / exp flow -
sensor
Reverse insp / exp Flow
Tv not achieved
System Leak
Unable to Drive Bellows
Volume sensors disagree
Alarms of Type:
Apnea; Apnea > 120s
No Fresh Gas Flow!
Unable to Drive Bellows
Ventilate Manually
Ventilator has no Drive Gas
Vol vent only. No PEEP or -
PSV
Ppeak Low. Leak?
Alarms of Type:
Inspiration Stopped
Negative airway pressure
PEEP High. Blockage?
Ppeak high
Symptoms of Type:
Continuously High Peep
during operation
Symptoms
Symptoms of Type:
No Pressure Waveform
No Flow Waveform
No Bellows Movement
Symptoms of Type:
Set T
V
vs. Measured T
V
Symptoms of Type:
Bellows Falls
Leak detected
No
Did System go to
Alt O2 or System
Malfunction?
Proceed to “Alt O2 / System
Malfunction Troubleshooting”
YesNo
Problem Solved
Legend:
ACB = Anesthesia Control Board
PCB = Power Controller Board
VIB = Ventilator Interface Board
DU = Display Unit
GIV = Gas Inlet Valve
APL = Adjustable Pressure Limit
(Flowchart 7)
(Flowchart 11)
(Flowchart 8)
(Flowchart 9)
(Flowchart 10)
(Section 7.9)
(Section 7.6)
AB.75.070

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet