Aisys
4-12 04/08 M1046983
4.3.1 Configuration
* For machines without a separate auxiliary common gas outlet and selector
switch, set Gas Outlet to SCGO: Selectable Common Gas Outlet.
* For machines with an external auxiliary common gas outlet and selector
switch, set Gas Outlet to ACGO: Auxiliary Common Gas Outlet.
Menu Item Message text Values Comments
Decimal Marker Select decimal
delineator.
0.01, 0 01 or 0,01
Language Change language
translation of screen
texts.
Chinese (simplified)
Czech
Danish
Dutch
English
Finnish
French
German
Greek
Hungarian
Italian
Japanese
Norwegian
Polish
Portuguese
Russian
Spanish
Swedish
Turkish
Default: English
Gas supply Colors Change color of O2,
N2O, and Air.
ANSI, ISO, Neutral ANSI: O2 green, Air yellow,
N2O blue;
ISO: O2 white, Air black/white,
N2O blue;
Neutral: All gases white.
O2 Flowtube O2 on left or right-
hand side.
Left, Right
Ventilator Drive Gas Change drive gas to
match machine
configuration.
Air, O2
Altitude Change altitude used
for gas calculations.
–400 to 3000 m
in 100-m increments
Gas Outlet* Change type of fresh
gas outlet.
SCGO, ACGO SCGO: Use insp port.
ACGO: Use auxiliary port.
N2O Enabled Change to match
machine
configuration.
Yes; No

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals