EasyManuals Logo

GE Datex-Ohmeda Aisys Carestation User Manual

GE Datex-Ohmeda Aisys Carestation
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #471 background imageLoading...
Page #471 background image
12 Service Application
M1046983 04/08 12-5
12.3.2 Gas Delivery Schematic
The Softkeys at the bottom of the Gas Delivery schematic bring up related diagnostic
screens that are also accessible from the Gas Delivery Subsystem menu.
• Gas Supplies — (Section 12.8.1)
• Mixer Output — (Section 12.8.2)
• Mixer Press/Temp — (Section 12.8.5)
Notes
Settings Remarks/Values
Flow/Concentration Move slider to increase/decrease value. Use arrow keys to
fine tune setting.
O2 Concentration
(if O2 Concentration selected)
0%, 21%, 25% to 100%.
21% minimum for Air Balance Gas
0% minimum for N2O Balance Gas
Total Flow
(if O2 Concentration selected)
Off, 0.10 l/min to 0.50 l/min (in 10 ml/min increments)
0.50 l/min to 15.0 l/min (in 0.25 l/min increments)
* O2 Flow
(if Individual Flow selected)
0.00 l/min to 15.00 l/min
* Balance Gas Flow
(if Individual Flow selected)
Air = 0.00 l/min to 15.00 l/min
N2O = 0.00 l/min to 12.00 l/min
* Total Fresh Gas Flow limited to 15.00 l/min

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters