7 Troubleshooting
M1046983 04/08 7-85
7.10.5 eVap Flow and Zero troubleshooting
Flowchart
20
Start
Is the Error entry
“Flow” related or
“Zero” related?
Flow Zero
Flow Related Error Log Entry
INFLOW CHECK VALVE FAILURE
OUTPUT FLOW LIMIT REACHED
CLOSED LOOP CONTROL FAILURE
Zero Related Error Log Entries
INFLOW ZERO POINT ERROR
OUTFLOW ZERO POINT ERROR
Perform the following tests:
• Backpressure Test
• Inflow Check Valve Test
Replace any components that fail the tests.
Within the PC Service Application,
verify the “zero” status
of the following fields:
• Inflow Flow
• Outflow Flow
Replace the Flowmeter Subassembly.
(Section 7.13)
(Section 7.12)
AB.75.177
(Section 12.10.1)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals