Aisys
7-6 04/08 M1046983
7.5 Troubleshooting System Malfunction (safe-state) screen
Machine logs • Compatibility incomplete: No versions received from Vent SIB
• System Self-tests failed
The above error log entries are due to the Ventilator failing its Power-On Self Test (Post).
When the ventilator fails its self-test, the system enters the safe-state.
The likely cause is a Gas Inlet Valve Solenoid that is not transitioning to the
de-energized state when valve drive power is removed.
Troubleshooting above stated condition:
• Replace the Gas Inlet Valve Solenoid (Section 9.13.2).
• If problem continues, replace the Gas Inlet Valve components (Section 9.13.3).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Datex-Ohmeda Aisys Carestation and is the answer not in the manual?

GE Datex-Ohmeda Aisys Carestation Specifications

General IconGeneral
TypeAnesthesia Machine
ManufacturerGE Healthcare
ModelAisys Carestation
Power Requirements100-240V AC, 50/60 Hz
DisplayHigh-resolution color touchscreen
Ventilator TypeElectronically controlled ventilator
Ventilation ModesVolume Control, Pressure Control, SIMV, Pressure Support, Manual
Gas DeliveryOxygen, Nitrous Oxide, Medical Air
Power SupplyAC mains, Battery backup
Integrated MonitoringYes
CommunicationEthernet
ConnectivityEthernet
AlarmsComprehensive alarm system for ventilation, gas delivery, and patient monitoring parameters

Related product manuals